QC Chart of Internal Quality Control (IQC)for 17-OHP Quantitative Test 

Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test


17-OHP IQC NS MIS System [login]
中文


Supported by Preventive Medicine Foundation

Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab


QC Chart of Internal Quality Control (IQC)for 17-OHP Quantitative Test

Select LotNo:
NS002


 CICII
QC Control Lot No.TPC1502ACTPC1502AD
ReagentKitPerkinElmer-GSP
Unitng/mL blood
Duration of the AnalyzingMonth
(2022/02)
CUM
(2018/04/01~2022/02/28)
Month
(2022/02)
CUM
(2018/04/01~2022/02/28)
Runs (N)4460344603
Mean7.210.035.960.1
SD0.801.683.5612.83
CV (%)11.116.89.921.3
Target Value12.112.166.866.8
Total Error (%)62.751.066.152.7
TEa (%)20202020
σ-1.80.2-2.70.5

 CICII
QC Control Lot No.TPC1502ACTPC1502AD
ReagentKitPerkinElmer-AutoDELFIA
Unitng/mL blood
Duration of the AnalyzingMonth
(2022/02)
CUM
(2018/04/01~2022/02/28)
Month
(2022/02)
CUM
(2018/04/01~2022/02/28)
Runs (N)06620662
Mean-10.8-66.6
SD-1.88-10.19
CV (%)-17.4-15.3
Target Value12.112.166.866.8
Total Error (%)-45.6-30.9
TEa (%)20202020
σ-0.5-1.3
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit: ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 12.1 ; SD : 1.60      ( ●)
 
Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit: ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 66.8 ; SD : 7.30      ( □)
NS002
SDI Diagram of CI & CII
Lot No.: TPC1502AC ; Reagent : PerkinElmer-AutoDELFIA ; Unit: ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 12.1 ; SD : 1.60      ( ●)
 
Lot No.: TPC1502AD ; Reagent : PerkinElmer-AutoDELFIA ; Unit: ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 66.8 ; SD : 7.30      ( □)
NS002
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: TPC1502AC ; Reagent : PerkinElmer-GSP ; Unit : ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 12.1 ; SD : 1.60
NS002
SDI Diagram of CI
Lot No.: TPC1502AC ; Reagent : PerkinElmer-AutoDELFIA ; Unit : ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 12.1 ; SD : 1.60
NS002
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: TPC1502AD ; Reagent : PerkinElmer-GSP ; Unit : ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 66.8 ; SD : 7.30
NS002
SDI Diagram of CII
Lot No.: TPC1502AD ; Reagent : PerkinElmer-AutoDELFIA ; Unit : ng/mL blood
Duration : 2018-04-01 to 2022-02-28 ; Target : 66.8 ; SD : 7.30
NS002
SDI Diagram of CII

Month :    ;    Cumulative : from to
[TOP]



Please select a Lab

NS001 NS002 NS003 NS004 NS005
NS006 NS007 NS008 NS009 NS010
NS011 NS012 NS013 NS014 NS015
NS016 NS017 NS018 NS019 NS020
NS021 NS022 NS026 NS031 NS032
NS034 NS035 NS999

[TOP]



Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab

Contact us
Address: P.O. Box 26-624 Taipei, Taiwan, 10699  Tel:+886-2-703-6080
    e-mail:
Fax: +886-2-2703-6070

Supported by:
      Preventive Medicine Foundation, Taipei, Taiwan
          updated : 29 Mar, 2022

[ Home ]