Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
17-OHP IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| | CI | CII |
| QC Control Lot No. | TPC1502AC | TPC1502AD |
| ReagentKit | PerkinElmer-DELFIA |
| Unit | ng/mL blood |
| Duration of the Analyzing | Month (2024/06) | CUM (2018/04/01~2024/06/30) | Month (2024/06) | CUM (2018/04/01~2024/06/30) |
| Runs (N) | 9 | 791 | 9 | 791 |
| Mean | 14.7 | 21.4 | 80.9 | 121.3 |
| SD | 1.04 | 8.32 | 7.70 | 53.96 |
| CV (%) | 7.1 | 38.9 | 9.5 | 44.5 |
| Target Value | 17.0 | 17.0 | 92.8 | 92.8 |
| Total Error (%) | 27.7 | 103.6 | 31.9 | 119.7 |
| TEa (%) | 13 | 13 | 13 | 13 |
| σ | -0.1 | -0.3 | 0.0 | -0.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC1502AC ; Reagent : PerkinElmer-DELFIA ; Unit: ng/mL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 17.0 ; SD : 2.21 ( ●) | |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-DELFIA ; Unit: ng/mL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 92.8 ; SD : 12.06 ( □) | |
 |
Lot No.: TPC1502AC ; Reagent : PerkinElmer-DELFIA ; Unit : ng/mL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 17.0 ; SD : 2.21 | |
 |
Lot No.: TPC1502AD ; Reagent : PerkinElmer-DELFIA ; Unit : ng/mL blood
Duration : 2018-04-01 to 2024-06-30 ; Target : 92.8 ; SD : 12.06 | |
 |