Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
17-OHP IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| | CI | CII |
| QC Control Lot No. | TPC2101AG | TPC2101AH |
| ReagentKit | PerkinElmer-GSP |
| Unit | ng/mL blood |
| Duration of the Analyzing | Month (2025/11) | CUM (2023/04/01~2025/11/30) | Month (2025/11) | CUM (2023/04/01~2025/11/30) |
| Runs (N) | 24 | 366 | 24 | 366 |
| Mean | 12.0 | 11.9 | 57.2 | 56.8 |
| SD | 1.37 | 1.66 | 35.06 | 17.76 |
| CV (%) | 11.4 | 13.9 | 61.3 | 31.3 |
| Target Value | 12.7 | 12.7 | 52.3 | 52.3 |
| Total Error (%) | 28.3 | 34.2 | 132.0 | 71.1 |
| TEa (%) | 20 | 20 | 20 | 20 |
| σ | 1.3 | 1.0 | 0.2 | 0.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit: ng/mL blood
Duration : 2023-04-01 to 2025-11-30 ; Target : 12.7 ; SD : 1.30 ( ●) | |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit: ng/mL blood
Duration : 2023-04-01 to 2025-11-30 ; Target : 52.3 ; SD : 2.47 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit : ng/mL blood
Duration : 2023-04-01 to 2025-11-30 ; Target : 12.7 ; SD : 1.30 | |
 |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit : ng/mL blood
Duration : 2023-04-01 to 2025-11-30 ; Target : 52.3 ; SD : 2.47 | |
 |
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022