17-OHP 缺乏症篩檢檢驗之實驗室內部品管計畫
Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
線上回報系統 (17-OHP IQC NS MIS System) [登入/login]
財團法人預防醫學基金會 執行
| | CI | CII |
| QC Control Lot No. | TPC2101AG | TPC2101AH |
| ReagentKit | PerkinElmer-GSP |
| Unit | ng/mL blood |
| 檢驗期間 | Month (2025/11) | CUM (2023/04/01~2025/11/30) | Month (2025/11) | CUM (2023/04/01~2025/11/30) |
| 累計點數 (N) | 24 | 366 | 24 | 366 |
| 累計平均值 (Mean) | 12.0 | 11.9 | 57.2 | 56.8 |
| 累計標準差 (SD) | 1.37 | 1.66 | 35.06 | 17.76 |
| 累計變異係數 (CV) % | 11.4 | 13.9 | 61.3 | 31.3 |
| 目標值 (Target) | 12.7 | 12.7 | 52.3 | 52.3 |
| 總誤差 (TE) % | 28.3 | 34.2 | 132.0 | 71.1 |
| 可容許總誤差 (TEa) % | 20 | 20 | 20 | 20 |
| σ | 1.3 | 1.0 | 0.2 | 0.4 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit: ng/mL blood
檢驗期間: 2023-04-01 至 2025-11-30 ; 目標值 (Target) : 12.7 ; 參考標準差 (SD) : 1.30 ( ●) | |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit: ng/mL blood
檢驗期間: 2023-04-01 至 2025-11-30 ; 目標值 (Target) : 52.3 ; 參考標準差 (SD) : 2.47 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit : ng/mL blood
檢驗期間: 2023-04-01 至 2025-11-30 ; 目標值 (Target) : 12.7 ; 參考標準差 (SD) : 1.30 | |
 |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit : ng/mL blood
檢驗期間: 2023-04-01 至 2025-11-30 ; 目標值 (Target) : 52.3 ; 參考標準差 (SD) : 2.47 | |
 |
若有任何疑問,歡迎來電(函)聯繫或指教
地 址:10699 台北郵政 26-624 號信箱 電 話:(02)2703-6080
e-mail:
傳 真:
(02)2703-6070
財團法人預防醫學基金會
updated : 29 Mar, 2022