Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
17-OHP IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| CI | CII |
QC Control Lot No. | TPC2101AG | TPC2101AH |
ReagentKit | PerkinElmer-GSP |
Unit | nmol/L blood |
Duration of the Analyzing | Month (2024/10) | CUM (2022/11/01~2024/10/31) | Month (2024/10) | CUM (2022/11/01~2024/10/31) |
Runs (N) | 30 | 630 | 30 | 630 |
Mean | 31.6 | 28.3 | 139.8 | 128.6 |
SD | 2.94 | 3.53 | 13.17 | 15.03 |
CV (%) | 9.3 | 12.5 | 9.4 | 11.7 |
Target Value | 29.5 | 29.5 | 136.1 | 136.1 |
Total Error (%) | 25.7 | 29.0 | 21.6 | 28.9 |
TEa (%) | 20 | 20 | 20 | 20 |
σ | 1.4 | 1.3 | 1.8 | 1.2 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit: nmol/L blood
Duration : 2022-11-01 to 2024-10-31 ; Target : 29.5 ; SD : 1.30 ( ●) | |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit: nmol/L blood
Duration : 2022-11-01 to 2024-10-31 ; Target : 136.1 ; SD : 2.47 ( □) | |
|
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit : nmol/L blood
Duration : 2022-11-01 to 2024-10-31 ; Target : 29.5 ; SD : 1.30 | |
|
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit : nmol/L blood
Duration : 2022-11-01 to 2024-10-31 ; Target : 136.1 ; SD : 2.47 | |
|
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022