Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
17-OHP IQC NS MIS System [login]
Supported by Preventive Medicine Foundation
| | CI | CII |
| QC Control Lot No. | TPC2101AG | TPC2101AH |
| ReagentKit | PerkinElmer-GSP |
| Unit | nmol/L blood |
| Duration of the Analyzing | Month (2026/01) | CUM (2022/11/01~2026/01/31) | Month (2026/01) | CUM (2022/11/01~2026/01/31) |
| Runs (N) | 30 | 930 | 30 | 930 |
| Mean | 29.1 | 28.2 | 121.4 | 129.1 |
| SD | 2.13 | 3.37 | 10.20 | 14.21 |
| CV (%) | 7.3 | 12.0 | 8.4 | 11.0 |
| Target Value | 29.5 | 29.5 | 136.1 | 136.1 |
| Total Error (%) | 16.0 | 28.3 | 27.6 | 27.2 |
| TEa (%) | 20 | 20 | 20 | 20 |
| σ | 2.6 | 1.3 | 1.1 | 1.4 |
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)
|
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit: nmol/L blood
Duration : 2022-11-01 to 2026-01-31 ; Target : 29.5 ; SD : 1.30 ( ●) | |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit: nmol/L blood
Duration : 2022-11-01 to 2026-01-31 ; Target : 136.1 ; SD : 2.47 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : PerkinElmer-GSP ; Unit : nmol/L blood
Duration : 2022-11-01 to 2026-01-31 ; Target : 29.5 ; SD : 1.30 | |
 |
Lot No.: TPC2101AH ; Reagent : PerkinElmer-GSP ; Unit : nmol/L blood
Duration : 2022-11-01 to 2026-01-31 ; Target : 136.1 ; SD : 2.47 | |
 |
Contact us
Address:
P.O. Box 26-624 Taipei, Taiwan, 10699 Tel:+886-2-703-6080
e-mail:
Fax:
+886-2-2703-6070
Supported by:
Preventive Medicine Foundation, Taipei, Taiwan
updated
: 29 Mar, 2022