QC Chart of Internal Quality Control (IQC)for 17-OHP Quantitative Test 

Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test


17-OHP IQC NS MIS System [login]
中文


Supported by Preventive Medicine Foundation

Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab


QC Chart of Internal Quality Control (IQC)for 17-OHP Quantitative Test

Select LotNo:
NS022


 CICII
QC Control Lot No.TPC1901AETPC1901AF
ReagentKitPerkinElmer-GSP
Unitnmol/L blood
Duration of the AnalyzingMonth
(2023/08)
CUM
(2020/06/01~2023/08/31)
Month
(2023/08)
CUM
(2020/06/01~2023/08/31)
Runs (N)1447714477
Mean18.516.990.584.5
SD1.231.499.056.98
CV (%)6.68.810.08.3
Target Value17.117.182.782.7
Total Error (%)21.518.829.418.7
TEa (%)20202020
σ1.82.11.12.1

 CICII
QC Control Lot No.TPC1901AETPC1901AF
ReagentKitPerkinElmer-AutoDELFIA
Unitnmol/L blood
Duration of the AnalyzingMonth
(2023/08)
CUM
(2020/06/01~2023/08/31)
Month
(2023/08)
CUM
(2020/06/01~2023/08/31)
Runs (N)02000200
Mean-15.9-87.7
SD-1.45-7.69
CV (%)-9.1-8.8
Target Value16.216.293.393.3
Total Error (%)-20.1-23.5
TEa (%)20202020
σ-2.0-1.6
Bias (%) = [ (│Mean - Target│) / Target ] x 100%
TE:Total Error(%) = Bias (%) + 2 × CV (%)
σ (Sigma) = [TEa% - Bias (%)] / CV (%)

Month :    ;    Cumulative : from to
[TOP]



CI and CII SDI
QC Chart

Lot No.: TPC1901AE ; Reagent : PerkinElmer-GSP ; Unit: nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 17.1 ; SD : 1.71      ( ●)
 
Lot No.: TPC1901AF ; Reagent : PerkinElmer-GSP ; Unit: nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 82.7 ; SD : 8.27      ( □)
NS022
SDI Diagram of CI & CII
Lot No.: TPC1901AE ; Reagent : PerkinElmer-AutoDELFIA ; Unit: nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 16.2 ; SD : 1.05      ( ●)
 
Lot No.: TPC1901AF ; Reagent : PerkinElmer-AutoDELFIA ; Unit: nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 93.3 ; SD : 5.69      ( □)
NS022
SDI Diagram of CI & CII

Month :    ;    Cumulative : from to
[TOP]



CI SDI QC Chart

Lot No.: TPC1901AE ; Reagent : PerkinElmer-GSP ; Unit : nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 17.1 ; SD : 1.71
NS022
SDI Diagram of CI
Lot No.: TPC1901AE ; Reagent : PerkinElmer-AutoDELFIA ; Unit : nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 16.2 ; SD : 1.05
NS022
SDI Diagram of CI

Month :    ;    Cumulative : from to
[TOP]



CII SDI QC Chart

Lot No.: TPC1901AF ; Reagent : PerkinElmer-GSP ; Unit : nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 82.7 ; SD : 8.27
NS022
SDI Diagram of CII
Lot No.: TPC1901AF ; Reagent : PerkinElmer-AutoDELFIA ; Unit : nmol/L blood
Duration : 2020-06-01 to 2023-08-31 ; Target : 93.3 ; SD : 5.69
NS022
SDI Diagram of CII

Month :    ;    Cumulative : from to
[TOP]



Please select a Lab

NS001 NS002 NS003 NS004 NS005
NS006 NS007 NS008 NS009 NS010
NS011 NS012 NS013 NS014 NS015
NS016 NS017 NS018 NS019 NS020
NS021 NS022 NS026 NS031 NS032
NS034 NS999

[TOP]



Home Application Peer Group Statistics (Table 1) Month vs. Cumulative (Table 2) QC Chart of Individual Lab

Contact us
Address: P.O. Box 26-624 Taipei, Taiwan, 10699  Tel:+886-2-703-6080
    e-mail:
Fax: +886-2-2703-6070

Supported by:
      Preventive Medicine Foundation, Taipei, Taiwan
          updated : 29 Mar, 2022

[ Home ]