17-OHP 缺乏症篩檢檢驗之實驗室內部品管計畫
Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
線上回報系統 (17-OHP IQC NS MIS System) [登入/login]
財團法人預防醫學基金會 執行
| | CI | CII |
| QC Control Lot No. | TPC2101AG | TPC2101AH |
| ReagentKit | Guangzhou Fenghua Bionegineering CO. LTD. |
| Unit | nmol/L blood |
| 檢驗期間 | Month (2025/11) | CUM (2024/07/01~2025/11/30) | Month (2025/11) | CUM (2024/07/01~2025/11/30) |
| 累計點數 (N) | 3 | 119 | 3 | 119 |
| 累計平均值 (Mean) | 39.0 | 47.3 | 134.4 | 154.1 |
| 累計標準差 (SD) | 5.87 | 5.09 | 2.23 | 13.51 |
| 累計變異係數 (CV) % | 15.1 | 10.8 | 1.7 | 8.8 |
| 目標值 (Target) | 42.0 | 42.0 | 133.6 | 133.6 |
| 總誤差 (TE) % | 37.2 | 34.1 | 3.9 | 32.9 |
| 可容許總誤差 (TEa) % | 20 | 20 | 20 | 20 |
| σ | 0.9 | 0.7 | >6 | 0.5 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: nmol/L blood
檢驗期間: 2024-07-01 至 2025-11-30 ; 目標值 (Target) : 42.0 ; 參考標準差 (SD) : 3.60 ( ●) | |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: nmol/L blood
檢驗期間: 2024-07-01 至 2025-11-30 ; 目標值 (Target) : 133.6 ; 參考標準差 (SD) : 12.20 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : nmol/L blood
檢驗期間: 2024-07-01 至 2025-11-30 ; 目標值 (Target) : 42.0 ; 參考標準差 (SD) : 3.60 | |
 |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : nmol/L blood
檢驗期間: 2024-07-01 至 2025-11-30 ; 目標值 (Target) : 133.6 ; 參考標準差 (SD) : 12.20 | |
 |
若有任何疑問,歡迎來電(函)聯繫或指教
地 址:10699 台北郵政 26-624 號信箱 電 話:(02)2703-6080
e-mail:
傳 真:
(02)2703-6070
財團法人預防醫學基金會
updated : 29 Mar, 2022