17-OHP 缺乏症篩檢檢驗之實驗室內部品管計畫
Internal Quality Control Program (IQC) for Neonatal 17-OHP Screening Test
線上回報系統 (17-OHP IQC NS MIS System) [登入/login]
財團法人預防醫學基金會 執行
| CI | CII |
QC Control Lot No. | TPC2101AG | TPC2101AH |
ReagentKit | Guangzhou Fenghua Bionegineering CO. LTD. |
Unit | nmol/L blood |
檢驗期間 | Month (2025/02) | CUM (2024/07/01~2025/02/28) | Month (2025/02) | CUM (2024/07/01~2025/02/28) |
累計點數 (N) | 10 | 103 | 10 | 103 |
累計平均值 (Mean) | 44.9 | 47.4 | 152.2 | 154.4 |
累計標準差 (SD) | 8.30 | 5.06 | 13.09 | 13.70 |
累計變異係數 (CV) % | 18.5 | 10.7 | 8.6 | 8.9 |
目標值 (Target) | 45.0 | 45.0 | 150.0 | 150.0 |
總誤差 (TE) % | 37.2 | 26.7 | 18.7 | 20.7 |
可容許總誤差 (TEa) % | 20 | 20 | 20 | 20 |
σ | 1.1 | 1.4 | 2.2 | 1.9 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: nmol/L blood
檢驗期間: 2024-07-01 至 2025-02-28 ; 目標值 (Target) : 45.0 ; 參考標準差 (SD) : 4.50 ( ●) | |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: nmol/L blood
檢驗期間: 2024-07-01 至 2025-02-28 ; 目標值 (Target) : 150.0 ; 參考標準差 (SD) : 15.00 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : nmol/L blood
檢驗期間: 2024-07-01 至 2025-02-28 ; 目標值 (Target) : 45.0 ; 參考標準差 (SD) : 4.50 | |
 |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : nmol/L blood
檢驗期間: 2024-07-01 至 2025-02-28 ; 目標值 (Target) : 150.0 ; 參考標準差 (SD) : 15.00 | |
 |
若有任何疑問,歡迎來電(函)聯繫或指教
地 址:10699 台北郵政 26-624 號信箱 電 話:(02)2703-6080
e-mail:
傳 真:
(02)2703-6070
財團法人預防醫學基金會
updated : 29 Mar, 2022