| | CI | CII |
| QC Control Lot No. | TPC2101AG | TPC2101AH |
| ReagentKit | Guangzhou Fenghua Bionegineering CO. LTD. |
| Unit | nmol/L blood |
| 檢驗期間 | Month (2026/01) | CUM (2024/07/01~2026/01/31) | Month (2026/01) | CUM (2024/07/01~2026/01/31) |
| 累計點數 (N) | 12 | 139 | 12 | 139 |
| 累計平均值 (Mean) | 41.1 | 46.5 | 137.1 | 151.2 |
| 累計標準差 (SD) | 2.13 | 5.21 | 11.74 | 14.91 |
| 累計變異係數 (CV) % | 5.2 | 11.2 | 8.6 | 9.9 |
| 目標值 (Target) | 41.0 | 41.0 | 126.0 | 126.0 |
| 總誤差 (TE) % | 10.6 | 35.8 | 25.9 | 39.7 |
| 可容許總誤差 (TEa) % | 20 | 20 | 20 | 20 |
| σ | 3.8 | 0.6 | 1.3 | 0.0 |
誤差 (Bias)% = { [ │累計平均值 (Mean) - 目標值 (Target)│] / 目標值 (Target) } x 100%
總誤差 (TE, Total Error)% = 誤差 (Bias)% + 2 × 累計變異係數 (CV)
σ (Sigma) = (TEa% - 誤差 (Bias)%) / 累計變異係數 (CV)
|
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: nmol/L blood
檢驗期間: 2024-07-01 至 2026-01-31 ; 目標值 (Target) : 41.0 ; 參考標準差 (SD) : 4.10 ( ●) | |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit: nmol/L blood
檢驗期間: 2024-07-01 至 2026-01-31 ; 目標值 (Target) : 126.0 ; 參考標準差 (SD) : 12.60 ( □) | |
 |
Lot No.: TPC2101AG ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : nmol/L blood
檢驗期間: 2024-07-01 至 2026-01-31 ; 目標值 (Target) : 41.0 ; 參考標準差 (SD) : 4.10 | |
 |
Lot No.: TPC2101AH ; Reagent : Guangzhou Fenghua Bionegineering CO. LTD. ; Unit : nmol/L blood
檢驗期間: 2024-07-01 至 2026-01-31 ; 目標值 (Target) : 126.0 ; 參考標準差 (SD) : 12.60 | |
 |